SCHMC

Comparison of PNA probe-based real-time PCR and Cobas TaqMan MTB for detection of MTBC

Metadata Downloads
Abstract
We compared the sensitivities and specificities of peptide nucleic acid (PNA) probe-based real-time PCR (PNAqPCR (TM) TB/NTM) and Cobas TaqMan MTB assays for detection of Mycobacterium tuberculosis complex (MTBC) in clinical specimens. A total of 425 clinical specimens including 360 respiratory specimens and 65 non-respiratory specimens were evaluated for comparative analysis. In respiratory specimens, the sensitivity of TaqMan MTB and PNAqPCR assay for detection of MTBC was 82.9% and 91.5%, respectively. In non-respiratory specimens, the sensitivity of the TaqMan MTB and PNAqPCR assay was 23.1% and 76.9%, respectively. Overall, the sensitivity and specificity of the TaqMan MTB assay for detection of MTBC was 76.9% and 100%, respectively. The PNAqPCR assay had a sensitivity and specificity of 90% and 99.7%, respectively.
All Author(s)
J. W. Kim ; Y. J. Choi ; H. J. Kim ; J. S. Park ; H. S. Nam ; Y. Hwangbo ; D. U. Kim ; K. S. Park
Issued Date
2013
Type
Article
Keyword
Peptide nucleic acidsPCRMycobacterium tuberculosis complexTuberculosisReal-time PCR
Publisher
한국바이오칩학회
ISSN
1976-0280
Citation Title
Biochip Journal
Citation Volume
7
Citation Number
2
Citation Start Page
85
Citation End Page
88
Language(ISO)
eng
DOI
10.1007/s13206-013-7201-0
URI
http://schca-ir.schmc.ac.kr/handle/2022.oak/1549
공개 및 라이선스
  • 공개 구분공개
파일 목록
  • 관련 파일이 존재하지 않습니다.